Chapter 6: Maintaining Your Enclosure

Product Service 1-800-522-7658

41

6. If the red light alarm is on, slowly turn the adjustment screw counterclockwise until

the green light is activated. If the green light is on, slowly turn the adjustment
screw clockwise until the red light comes on. Slowly turn the adjustment screw
back until the red light is activated. It is important that these adjustments be done
in small increments, at intervals about 30 seconds apart to allow for delayed
reaction of the alarm itself. The alarm low flow set point should now be set and the
red light activated.

7. Readjust the enclosure airflow to its normal operating levels. The green light

should now be activated. Calibration is now complete.

Enclosure Operating In Flow Speed

Alarm Condition Set Point Speed

100 fpm

80 - 90 fpm

80 fpm

60 - 70 fpm

60 fpm

40 - 50 fpm

8. Note: If the low airflow volume or sash opening cannot be adjusted, then a 1/4 to 1/3

of a turn counterclockwise can be adjusted to set the airflow alarm condition at 10-
25% below normal operating levels.

Guardian 500 Alarm Activation

The audio and visual alarm will activate approximately 10-30 seconds after an alarm condition is
detected. To temporarily mute the audible alarm, press and release the test/reset button.

NOTE: After an alarm condition has been detected, the red light will stay on. The
audible alarm will remain muted until airflow returns to normal levels.