Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

26

Information Center

The Information Center is an LCD display located on the right side wall at
eye level. The center’s display is divided into three separate areas, as shown
in Figure 5-2, below.

Status Area

Data Area

Icon Area

Status Area

The Status Area displays the system’s condition during operation. This line
can display any of the following messages:
OK

The motor is operating properly, and the sash is not open too far.


Sash is too high

The sash is open too far for safe operation.


Airflow Alert

The airflow patterns in the cabinet have changed, resulting in a sudden
change in the motor speed. This is most likely due to a blockage of the
grille or the exhaust filter outlet. It may also be caused by removal of the
work surface while the cabinet is in operation.


System Error

The motor and display circuit board are not communicating properly.
DO NOT USE THE CABINET UNTIL THE PROBLEM HAS BEEN
CORRECTED.

Figure 5-2