Labconco PuriCare Procedure Stations 38100 User Manual

Page 48

Advertising
background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

43

Wipe-Down

• Raise the sash to its full open position (approximately 21.75 inches/

55.25 cm). Mute the alarm by depressing the "Alarm Silence" switch.

• Wipe down the interior surfaces of the cabinet with 70% ethanol, or a

suitable disinfectant, and allow to dry.

Loading Materials and Equipment

• Only load the materials required for the procedure. Do not overload

the cabinet.

• Do not obstruct the front, side, or rear return air grilles.

• Large objects should not be placed close together.

• Slowly close the sash until it is in the correct operating position.

• After loading the cabinet, wait two to three minutes to purge airborne

contaminants from the work area.

Work Techniques

• Keep all materials at least 4 inches (10.16 cm) inside from the sash,

and perform all contaminated operations as far to the rear of the work
area as possible.

• Segregate all clean and contaminated materials in the work area.

• Arrange materials to minimize the movement of contaminated

materials into clean areas.

• Keep all discarded contaminated material to the rear of the work area.

• Avoid moving materials or the operator's hands and arms through the

front access opening during use.

• Avoid the use of an open flame. Use disposable lab ware or an

electric incinerator as alternatives.

• Use proper aseptic technique.
• Avoid using techniques or procedures that disrupt the airflow patterns

of the cabinet.

• If there is a spill or splatter during use, all objects in the cabinet

should be surface decontaminated before removal. Thoroughly
disinfect the working area of the cabinet WHILE IT IS STILL IN
OPERATION. This will prevent the release of contaminants from the
cabinet.

Advertising
This manual is related to the following products: