Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

42

If An Airflow Alert Activates

The most common causes of an airflow alert are:

• Blockage of the inlet grilles or exhaust outlet.

• Removal of the work surface or grille during operation.

Note: If the alarm sounds during use, immediately take appropriate action to
prevent contamination to you and other personnel in the area.

Resetting the Airflow Alert System

The Airflow Alert will automatically reset to normal operation once the
motor speed has stabilized.

Working In the Procedure Station

Planning

• Thoroughly understand procedures and equipment required before

beginning work.

• Arrange for minimal disruptions, such as room traffic or entry into the

room while the cabinet is in use.

Start-up

• Turn off UV light if included.

• Slowly raise the sash until the bottom of the sash aligns with the sash

indicator decal located on the left side of the work area.

• Turn on fluorescent light and cabinet blower if the SmartStart

features

have not been activated.

• Check the air grilles for obstructions, and note the filter gauge

reading.

• Allow the cabinet to operate unobstructed for 5 minutes.

• Wash hands and arms thoroughly with germicidal soap.

• Wear a long sleeved lab coat with knit cuffs and over-the-cuff rubber

gloves. Use protective eyewear. Wear a protective mask if
appropriate.