Uv lamp timer, Uv lamp maintenance – Labconco PuriCare Procedure Stations 38100 User Manual

Page 42

Advertising
background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

37

UV Lamp Timer

The UV Lamp Timer allows you to decide whether the UV lamp will operate
continuously or for a time period that you select.

Interval

When the “Interval” option is selected, the UV Interval Timer menu is
displayed.


UV Interval Timer

This window lets you set the amount of time the UV lamp
stays lit when activated. Use the “▲” and “▼” buttons to
select the amount of time you need, then press the “OK”
button to set the interval.

NOTE: Raising the sash will reset the timer to its original
setting. Pressing “UV Light” will turn the light off,
pressing “Mute/OK“ will pause the timer.

Continuous

In the continuous mode, the UV light will remain lit until the “UV Light”
button is pressed, or the sash is raised.

UV Lamp Maintenance

This selection allows you to check how many hours the UV lamp has burned.
It also allows you to set the number of hours you want the lamp to last, and to
reset the UV lamp hourmeter.

UV Lamp Hourmeter

This display only shows how many hours the UV lamp has been lit, and
how many hours remain until you will receive a warning to replace the
lamp.

Change UV Lamp Life

In this option, you can set the number of operating hours before
receiving the replace UV lamp warning. For most UV lamps, the output
of UV light decreases at a constant rate. Typically, after 6,000 hours of
operation the lamp will output 80% of the UV light it did when it was
new. This option allows you to set the warning at a life you prefer, in
100 hour increments.

Reset UV Lamp Clock

This option lets you reset the UV hourmeter to 0 whenever the lamp has
been replaced.