Chapter 2: Prerequisites

Product Service 1-800-522-7658

3

Location Requirements

Note: The Procedure Station should be located away from traffic patterns,
doors, fans, ventilation registers, fume hoods and any other air-handling
devices that could disrupt its airflow patterns. All windows in the room
should remain closed. Figure 2-1 shows the preferred location for the
Procedure Station. The preferred location is perpendicular to the door, to
minimize the effect of opening and closing the door on the performance of
the cabinet.


Preferred location

Air register blocked or redirected

to prevent cabinet disruption

Alternate location










Figure 2-1