Chapter 7: Troubleshooting

Product Service 1-800-522-7658

53

PROBLEM

CAUSE

CORRECTIVE ACTION

Contamination of
work in the cabinet

Improper technique
or procedure for the
cabinet

See “Use of the Cabinet” section in the
manual.

Restriction of the
return air slots or
grille – blockage of
the exhaust outlet

Ensure that all return air slots, grilles
and the exhaust outlet are unobstructed.

External factors are
disrupting the cabinet
airflow patterns or
acting as a source of
contamination

See “Installation” section of this
manual.

Cabinet is out of
adjustment/HEPA
filter(s) are defective

Have cabinet recertified.