The common service operations necessary to maintain the Procedure Station
for peak performance are listed below.

Routine Maintenance Schedule

Weekly

• Using 70% ethanol, or a suitable disinfectant, surface disinfect the

inside of the cabinet and the work surface.

• Using an appropriate glass cleaner, or LabSolutions Glass & Surface

Wipes, Labconco part # 1570000; clean the sash and the surface of
the UV lamp, if so equipped.

• Operate the cabinet blower, noting the percent filter life remaining in

an operational log.

Monthly (or more often as required)

• Using a damp cloth, or LabSolutions Glass & Surface Wipes,

Labconco part # 1570000; clean the exterior surfaces of the cabinet,
particularly the front and top of the cabinet, to remove any
accumulated dust.

• Disinfect and lift the work surface. Surface disinfect the lower

plenum with a solution of 70% ethanol, or a suitable disinfectant.
Check the towel catch and prefilter for retained materials. Replace
prefilter if necessary.

• Check all service valves, if so equipped, for proper operation.
• Check the UV and fluorescent light hourmeters, and record their

readings in an operational log.

• All weekly activities.

Semiannually or Annually

• Have the cabinet re-certified by a qualified certification technician.
• All monthly activities.