Labconco PuriCare Procedure Stations 38100 User Manual

Page 32

Advertising
background image

Chapter 5: Using The Cabinet

Product Service 1-800-522-7658

27

Data Area

The Data Area displays the Filter Gauge. This bar graph directly indicates the
percent of remaining filter life. When the cabinet is new, or after the HEPA
filters are replaced, the gauge will be “full”, or display approximately 100%.
As the filters load, or if the grille or exhaust discharge are blocked, the bar
graph will show a reduced filter life.

The graph is user-selectable as a bar graph with percentage displayed, a bar
graph only, or a semi-circular graph showing filter life remaining.

When the optional airflow sensor is installed and calibrated, the inflow and
calculated downflow velocities are also displayed during operation.

Icon Area

The Icon Area displays icons for those features selected or in operation. The
Icons are:

Normal blower operation –

In this mode the cabinet blower is operating normally.

NightSmart

TM

blower operation

When NightSmart

is selected, the blower will slow to idle speed

when the sash is closed, to maintain the cleanliness of the air in
the work area. When the sash is opened, the blower will resume
normal operation.


Fluorescent light

In this mode the fluorescent lights are operating. When the sash is
fully closed, the fluorescent lights will automatically turn off.

UV light

In this mode the UV (germicidal) light is operating. When the
sash is fully closed, the UV light can be programmed to
automatically turn on. When the sash is raised, the light will
automatically turn off, to protect the laboratory from UV
irradiation.

Timer

When either the interval or stopwatch timer mode is active, this
icon appears.