Chapter 6: Maintaining The Cabinet

Product Service 1-800-522-7658

48

Postfilter Replacement:

If odors become noticeable from the exhaust of the cabinet, replace the
carbon postfilter on top of the cabinet. To do so, carefully compress the
perforated exhaust filter cover to pull the tabs out of one side of the exhaust
filter brackets. Set the cover aside. Remove and dispose the old postfilter.

Note: When working around the exhaust HEPA filter, be careful to
never touch the filter with hands or tools. If damaged, a qualified
certifier must repair or replace the HEPA filter.

Carefully place a new postfilter between the exhaust filter brackets with the
airflow arrows pointing up. Reinstall the perforated exhaust filter cover by
inserting the tabs into the slots of the brackets.

Front Panel Removal and Installation:





1. Locate and remove the two Phillips

screws that secure the front panel as
shown in Figure 6-3. They are
located on the bottom corners of the
front dress panel.

2. Swing the bottom of the dress panel

out to clear the fluorescent light and
then lift the front dress panel
straight up and away from the
cabinet.

Figure 6-3

Advertising