Chapter 6: Maintaining The Cabinet

Product Service 1-800-522-7658

47

Towel Catch and Prefilter Removal:

The towel catch should be removed periodically for inspection.

Note: The work surface and the towel catch must be thoroughly
decontaminated before removing either from the cabinet.

1. Remove the work surface as described above.

2. Lift the front end of the center support straight up to release it from the

locating pins (A). Set the support aside in the work area.

3. Remove the towel catch by pivoting the bottom out toward you (B).

4. Inspect the roughing prefilter. Disinfect, clean or replace as necessary.

Surface decontaminate the towel catch before removing it.

5. Reinstall the towel catch by sliding it back into position, ensuring that the

two slots on either end of the towel catch properly engage the rails on the
rear of the duct, as shown in Figure 6-2. Also ensure that the DOP
Sampling Line (the clear tube with a black cap) either passes through a
hole in the towel catch, or passes along the outside edge of the catch
without being kinked.

Figure 6-2