Calibration

Sorensen Ethernet Option

4-6

M130/M131 Programming Manual

12.

Program the supply to store the calibration values in non-volatile memory:

CAL[n]:STORE

Allow 10 seconds for the non-volatile memory to be updated with the
new calibration values.


13.

Program the supply to lock the non-volatile memory for calibration value
protection

CAL[n]:LOCK


14.

The voltage readback calibration is complete. The unit may be turned off or
other calibration procedures may be performed.

4.5 O

VERVOLTAGE

P

ROTECTION

P

ROGRAMMING

The overvoltage protection calibration procedure requires calibrated output voltage
programming and voltage readback.

STEP

DESCRIPTION

1.

Program the supply to self-calibrate the overvoltage protection:

CAL[n]:OUTP:VOLT:PROT:CALC

The overvoltage protection calibration function requires over 30 seconds.


2.

Program the supply to unlock the non-volatile memory for calibration value
storage:

CAL[n]:UNLOCK “6867”


3.

Program the supply to store the calibration values in non-volatile memory:

CAL[n]:STORE

Allow 10 seconds for the non-volatile memory to be updated with the
new calibration values.


4.

Program the supply to lock the non-volatile memory for calibration value
protection

CAL[n]:LOCK


5.

The output overvoltage protection calibration is complete. The unit may be
turned OFF or other calibration procedures may be performed.