Seaward 300 PrimeTest User Manual
Page 81
PrimeTest 300
Operating Instructions
- 80 -
14.0 Electrical Specifications
14.1 Earth Continuity
Test Current (2Ω load)
±200 mA
Test Circuit Voltage
± 4V min
Test Duration
2, 5, 10, 20, 30 seconds
Pass/Fail Limit
User Defined
Display Range
0.00 – 19.99Ω
Measuring Range
0.01 – 19.99Ω
Resolution
0.01Ω
Accuracy
± 5% ±2 digits
14.2 Insulation Resistance
Test Voltage Open Circuit
500V +20% / -0%
Test Voltage @ 1mA
500V min
Test Current Short Circuit
<2 mA
Test Duration
2, 5, 10, 20, 30 seconds
Pass/Fail Limit
User Defined
Display Ranges
0.00 – 1.99MΩ
2.0 – 19.9MΩ
20 – 199MΩ
Measuring Range
0.01 – 40MΩ
Resolution
0.01MΩ, 0.00 – 1.99MΩ
0.1MΩ,
2.0 – 19.9MΩ
1MΩ, 20 – 199MΩ
Accuracy
± 5% ±2 digits
14.3 IEC Lead Test
This is a 3 part test is made up of:
Earth Continuity – specification as 10.1 above.
Insulation Resistance – specification as 10.2 above.
Polarity Check – checks for: Live open
Neutral open
L/N Short
L/N reversed