Phase angle (f) (calculation out of power factor)

Display of phase angle of an individual input or the total value, as desired
Measurement range: .... 0° to 90°
A negative sign “- “shows a leading current.
Resolution: ..................0.1°
Accuracy: ....................± 4°

Frequency measurement (for voltage ramge > 80 V)

Measurement range: ....45 to 80 Hz
Resolution: ..................0.1 Hz
Accuracy: ....................± (1 % rdg + 10D) for voltage > 80 V

Rotary field detection (for conductor voltages > 80 V)

Display for clockwise rotary field:

, Display for counterclockwise rotary field:

Energy measurement – active power (kWh)
Display of the total value of all active power values

Range

Resolution

Accuracy

Sampling Rate

3.999 kWh

0.001 kWh

± (3.5% rdg. + 20D)

1 second

39.99 kWh

0.01 kWh

399.9 kWh

0.1 kWh

3.999 MWh

0.001 MWh

39.99 MWh

0.01 MWh

119.3 MWh

0.1 MWh

Energy measurement – apparent power (kVAh)
Display of the total value of all apparent power values.

Range

Resolution

Accuracy

Sampling Rate

3.999 kVAh

0.001 kVAh

± (3.5% rdg. + 20D)

1 second

39.99 kVAh

0.01 kVAh

399.9 kVAh

0.1 kVAh

3.999 MVAh

0.001 MVAh

39.99 MVAh

0.01 MVAh

119.3 MVAh

0.1 MVAh

Energy measurement – idle power ( kvarh)
Display of total values of all idle power values.

Range

Resolution

Accuracy

Sampling Rate

3.999 kvarh

0.001 kvarh

± (3.5% rdg. + 20D)

1 second

39.99 kvarh

0.01 kvarh

399.9 kvarh

0.1 kvarh

3.999 Mvarh

0.001 Mvarh

39.99 Mvarh

0.01 Mvarh

119.3 Mvarh

0.1 Mvarh

Harmonics measurement
This measurement is only possible via a PC.
Measurement up to the 31st harmonics
Measurement inputs: U1, U2, U3, I1, I2, I3
Voltage range > 80 V
Current range > 50 A
Sampling range: approx. 3 s.
Sampling resolution per measurement: 64 points