Appendix e, Terescopes bench test procedure – MRV Communications TERESCOPE TS800/155 User Manual

Page 65

Advertising
background image

M R V C o m m u n i c a t i o n s , I n c . – I n s t a l l a t i o n M a n u a l

55

TereScopes Bench Test Procedure


Introduction

All TS Products are bench tested indoors prior to outdoor

installation to ensure that the system is fully functional.
The bench test is a simple procedure whereby a link pair is aligned

on the table and activated to simulate a channel of communication
(see fig.1).

2 Points to Remember

1. Since the link distance during the bench test is very short (i.e. the

devices activated are very close), the receivers will go into saturation
unless the signal is attenuated. (NOTE: In the 700/G, deep into
saturation the DVM reading may actually drop back down, giving
the false impression that the link is misaligned. It is therefore
vital to place the attenuator in the link path before optimizing
alignment.)
To avert entering saturation, the transmit signal must be physically
attenuated.

We recommend the simple procedure of inserting a piece of paper
or the like into the beam path, or concealing a portion of the beam

with an opaque (non-transparent) material. This will reduce the
signal power entering the receiver.

Make sure to attenuate the signal enough so that the receiver’s
optical power meter value falls below the saturation estimate of the

device. See table below for saturation estimate.

2. An additional derivative of the short link distance is the presence
of reflections.

The signal will reflect off the front window of the receiver back at
the transmitting device and may be mistaken as part of the opposite

transmission.
This interference is commonly called “cross talk”.

To avoid cross talk during the bench test, it is advisable to check
whether interfering reflections exist by shutting off power to one

device and verifying that the optical power meter reading in the
other (active) device is zero.

This should be repeated for the opposite device.
Alternatively, a practical setup for bench testing the 4” series

(models B, C and D) and Light series (models A and C2) is
presented in Figure 1; the bench test setup for the 10” series

(models E and F) is presented in Figures 2a,2b.
In the 4”/Light setup, a thin physical barrier, such as a piece of

cardboard, is used as a wall to divide between the beam paths, thus
ensuring that no cross talk occurs.

In the 10” setup, the two devices are not centrally aligned; instead,
only one corner of each device faces the opposite device. This

APPENDIX E

Advertising
This manual is related to the following products: