Section 4. Electric Checks and Adjustments

SATO CL408e/CL412e Service Manual

PN 9001078

Rev. B

4.6 Label Gap Adjustment

Additional equipment required: TP Test Module

Digital Multimeter

STEP

PROCEDURE

Refer to Section 4.2 for access to main PCB

1.Turn VR4 (GAP) potentiometer on the main PCB all the way to the

left.

2.Refer to Section 4.

3.

Set the digital multimeter to DC voltage measurement mode.

Attach the connector from the TP Test Module to the test port on the

main PCB. Note correct positioning of connector. Nibs on the connector

are placed down in the receptacle on the PCB in the forward position.

Set the dial to 5.

3.Connect (+) probe of the multimeter to Sig+ and (-) probe to pin

GND.

4.For Low level (Label Gap part with backing only) adjustment, put label

gap part in the sensor. Then adjust the electrical level with VR4 on the

main PCB so that it will measure less than 0.5 V.

5.For High level (paper part) adjustment, put paper part in the sensor

and check the electrical level. If the level difference is +1.0 V more

than the Low level, it is acceptable. If it is lower than 1.0V repeat

STEPS 4 & 5 and readjust VR4.

6.Standard values:

Low level (gap): below 0.

5 V

High level (paper part): Low level +1.0 V or

higher. If these values do not result, try the

following:
a) Repeat the process
b) Clean the sensor
c) Verify sensor is operational
d) Replace labels with higher quality labels
e) Perform factory reset