2 calibration message, Procedure – Ohaus NBF1A2 Navigator Balances User Manual

Page 15

Advertising
background image

15

3.3.2 Calibration Message

On Navigator

TM

balances equipped with the InCal

TM

feature, a screen prompt of CAL NOW appears when the balance

requires calibration. This display can be turned off if it is desired not to have the balance indicate that calibration is
required. Turning the display off has no effect on the basic balance operation.

Procedure

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYP is displayed.

• Press

button, CAL NOW is displayed.

• Press Enter button, CAL NOW ON or OFF is dis-

played.

• Press

or

button and select either ON or

OFF. When OFF is selected, the CAL NOW message

will not appear in the display.

• Press Enter button, SAVED is momentarily displayed,

then the display indicates CAL ADJ.

NOTE: At this point you may continue with the calibra-
tion adjust procedure on the next page or exit.

The calibration adjust procedure is only used when it is

desired to calibrate the internal calibration mass to a
known external mass if a difference exits between the
known mass and the InCal

TM

value.

• Press Enter button, LOCK is displayed.

• Press

or

button and select either ON or OFF.

• Press Enter button, SAVED is momentarily displayed,

then the display indicates EXIT.

• Press Enter button, display returns to WEIGH mode.

Advertising