Calibration, 1 span calibration, Procedure – Ohaus NBF1A2 Navigator Balances User Manual

Page 71: Zero tare

Advertising
background image

8

TM

®

Print

Mode

Setup

Units

TM

Enter

On/Off

G/N/T

Zero

Tare

®

Print

Mode

Setup

Units

TM

Enter

On/Off

G/N/T

Zero

Tare

3.3.1 Span Calibration

Span calibration utilizes two calibration points, zero and full capacity. All balances will prompt for a particular weight.
Sample display below illustrates a 610g balance.

Procedure

• Clear the pan.

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYP is displayed.

• Press Enter button to select SPAN calibration, CAL

TYP SPAN is displayed.

• Press Enter button, BUSY is displayed.

NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.

• Place the calibration mass on pan.

NOTE: The PUT WEIGHT message indicates the
calibration mass that is on the pan.

• Press Enter button, BUSY is displayed. After a

few seconds CAL SET is displayed, the display then
returns to WEIGH mode.

• Span calibration is completed.

• Remove calibration mass from the pan.

Setup

Use keys to change, press Enter

Change

Setup

Use keys to change, press Enter

Change

Setup

Change

• Display changes to PUT WT 600g. The displayed

weight is the value of the mass to be used.

Setup

press Enter

g

Change

Setup

Use keys to change, press Enter

Change

Stable
Center
of Zero

g

B/G

Advertising