3 calibration adjust, Procedure – Ohaus NBF1A2 Navigator Balances User Manual

Page 16

Advertising
background image

16

3.3.3 Calibration Adjust

Balances with InCal

TM

contain software which allows the

internal calibration value to be adjusted by + 100 divisions
at full scale capacity using an external mass which is
traceable to a certified standard.

Procedure

• Perform the internal calibration procedure.

• Press >O/T< button to zero the balance.

• Place a mass equal to the

span calibration value of

the balance on the pan. Note the reading on the bal-

ance and see if the balance indicates the exact

weight or indicates a higher or lower reading. If the

reading is higher or lower, proceed.

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYP is displayed.

• Press

button until , CAL ADJ is displayed.

• Press Enter button, CAL ADJ 0 should be dis-

played (0 is factory setting).

NOTE: Balance will retain last CAL ADJ setting.

• Press

or

button until desired number is

displayed.

• Press Enter button, SAVED is momentarily displayed,

then display indicates LOCK.

• Press Enter button, LOCK ON or LOCK OFF is

displayed.

• Press

or

button and select either LOCK ON

or LOCK OFF.

• Press Enter button, SAVED is momentarily displayed,

then EXIT is displayed.

• Press Enter button, display returns to WEIGH

mode. Note weight reading and remove mass from

pan.

• Perform the internal calibration procedure. The value

entered as an adjustment is now stored. Place the

calibration mass on the pan and check. Repeat pro-

cedure if further correction is required.

• To return to factory setting, follow procedure above

and set CAL ADJ to 0.