5 user calibration, Procedure – Ohaus NBF1A2 Navigator Balances User Manual

Page 18

Advertising
background image

18

3.3.5 User Calibration

User calibration is used when it is desired to calibrate the balance using a mass of known value. User calibration limits
are no less than the 1/2 nominal capacity point and no more than the display over limit. For example, on a 210g
balance, it's no less than 100.00g and no more than 210.09g. To use this calibration feature, proceed as follows:

(FLASHING)

• Clear the pan.

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYP is displayed.

• Press Enter button.

• Press

and scroll to select USER calibration,

CAL TYP USEr is displayed.

• Press Enter button, the display indicates the last

calibration mass value which was entered with the

first digit flashing. (Sample illustrates 200g).

• Press

to change value of digit and

button to advance to next digit. Enter the desired
mass value for your balance. This number must be
at least 50% of the full span value.

• After entering value, press Enter button, BUSY is

displayed.

NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.

• Display changes to PUT WT 200 g.

• Place specified calibration mass on pan.

• Press Enter button, BUSY is displayed. After a

few seconds CAL SET is displayed, the display then
returns to WEIGH mode.

• User calibration is completed.

• Remove calibration mass from the pan.