5.3 Error Codes List

Error Codes List

The following list describes the various error codes which can appear on the display and the suggested remedy.

Data Errors

1.0

Transient error (hardware error, probably static discharge). If error persists, the balance must be
serviced.

1.1

Balance temperature transducer hardware error.

Tare Errors

2.0

Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needs
recalibration.

Calibration Errors

3.0

Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.

RS232 Errors

4.4

RS232 buffer is full.

User Errors

7.0

User entry out of bounds.

7.2

Number outside of display capacity.

Over-Under Load Errors

8.0

Hardware error causing an internal weight signal which is too low. Check if pan is off. If not, the balance
must be serviced.

8.1

Hardware error caused by an internal weight signal which is too high. Check load on the pan which may
be excessive. If error persists, the balance must be serviced.

8.2

Load on pan exceeds +10% power on zero limit.

8.3

Rated capacity exceeded. Remove excessive weight from pan.

8.4

Underload condition on balance. Check that the proper pan is installed.

CheckSum Errors

9.1

Bad factory checksum. If error persists, have the balance serviced.

9.2

Bad factory checksum. If error persists, have the balance serviced.

9.3

Bad factory checksum. If error persists, have the balance serviced.

9.5

Factory calibration data failed checksum.

9.8

User calibration data failed checksum.

9.9

Factory temperature compensation data failed checksum.

5.4 Information Messages

SAVED

This message is flashed when an item is changed in the menu and the new value is written
to the EEPROM.

LOCKED

This message is flashed when an item can not be changed in the menu because the menu
is locked and the Lock Switch is set locked.

LOW REF

The message is flashed in parts counting or percent when the calculated reference weight
is very low.

HI REF

This message is flashed when capacity is exceeded in % mode.

T ENT

This message is displayed when ZERO key ia pressed with a Tare value entered. Indicates
that zero command is ignored.

T SET

Displayed when Tare key is used to set a tare value.