3.6 Parts Counting

Parts Counting is

enabled only when Count is turned ON in the Mode menu under Setup. In the parts counting

mode, the balance displays the quantity of parts you place on the pan. Since the balance determines the quantity
based on the average weight of a single part, all parts must be reasonably uniform in weight. Refer to paragraph
4.7 to enable parts counting.

Procedure

• Press the Mode button.

• Press

button until COUNT is displayed.

• Press Enter button. PUTPAN 10 is displayed

(default setting). Balance will retain last sample size

saved.

• Put a container on the pan and press

to tare.

• Press and hold

or

button and select sample

size. Sample size may be 5 to 1000 parts.

• Place sample on the pan.

• Press Enter button to continue, display indicates

BUSY.

Balance displays the reference weight of an individual

part for five seconds and then displays the total number

of parts on the pan.

• Place parts to be counted on the pan. Balance dis-

plays number of parts.

• To view the total weight of the parts, press the UNITS

button, then press

to select desired weighing

unit, then press the Enter button.

• To return to parts counting, press the UNITS button,

then press

button until COUNT is displayed,

then press the Enter button.

• To begin counting with a new sample, repeat proce-

dure.

Update

Update is a function which permits placing additional
samples which are greater than the value of the original
sample but less than three times the value. This action
increases the accuracy of the measurement.

• Place sample on the pan which is at least one but not

more than three times the original sample size.

• Press Mode button, COUNT is displayed.
• Press Enter button, UPDATE is displayed.
• Press Enter button, BUSY is displayed then the

reference weight followed by the new sample size.