Rs232 interface & commands, 2 rs232 interface, Hardware – Ohaus NBF1A2 Navigator Balances User Manual

Page 42: Output formats, Rs232 commands

Advertising
background image

42

Command

Character

Description

?

Print current mode

nnnnA

Set Auto Print feature to “nnnn”
(see table).

RS232 COMMAND TABLE

Field: Mode Stab CR

LF

Length:

5

1

1

1

blank if stable
“ ? ” if unstable

nnnn = 0

Turns feature OFF

nnnn = S

Output on stability

nnnn = C

Output is continuous

nnnn = 1-3600 Sets Auto Print

Interval

5.2 RS232 Interface

Navigator

TM

balances are equipped with a bi-directional

RS232 compatible interface for communication with
printers and computers. When the balance is connected
directly to a printer, displayed data can be output at any
time by simply pressing PRINT, or by using the Auto Print
feature.

Connecting the balance to a computer enables you to
operate the balance from the computer, as well as receive
data such as displayed weight, weighing mode, stability
status, etc.

The following sections describe the hardware and soft-
ware provided with the balance

Hardware

On the rear of the balance, the 9-pin male subminiature
“D” connector is provided for interfacing to other devices.
The pinout and pin connections are shown in the adjacent
illustration.

The balance will not output any data unless pin 5 (CTS) is
held in an ON state (+3 to +15 V dc). Interfaces not
utilizing the CTS handshake may tie pin 5 to pin 6 to
defeat it.

IMPORTANT:
An Ohaus RS232 cable is required to operate with this
balance. A standard RS232 cable will not work. See
Accessories listed in paragraph 5.7.

Output Formats

Data output can be initiated in one of three ways: 1) By
pressing PRINT; 2) Using the Auto Print feature; 3)
Sending a print command (“P”) from a computer.

The output format is illustrated in the RS232 command
table which follows.

RS232 Connection

1

N/C

2

Data Out (TXD)

3

Data In (RXD)

4*

Tare (External signal)

5

Clear To Send (CTS)

6

Data Terminal Ready (DTR)

7

Ground

8

Request To Send (RTS)

9*

Print (External signal)

* External PRINT and/or TARE

switches may be installed
as shown in the diagram.
Momentary contact
switches must be used.

PRINT *

TARE *

RS232 Commands

All communication is accomplished using standard ASCII
format. Characters shown in the following table are
acknowledged by the balance. Invalid command re-
sponse "ES" error indicates the balance has not recog-
nized the command. Commands sent to the balance
must be terminated with a carriage return (CR) or carriage
return-line line feed (CRLF). Data output by the balance is
always terminated with a carriage return - line feed
(CRLF).

Rear of Balance

mg
g
kg
dwt
ct
oz
oz t

GN
tael HK (Hong Kong)
tael SG (Singapore)
tael RC (Taiwan)
momme
lb

N

tical
tola
custm
Pcs
%

Advertising