3 calibration glp printout, Span calibration printout, Linearity calibration printout – Ohaus NBF1A2 Navigator Balances User Manual
Page 73
10
TM
3.3.3 Calibration GLP Printout
If any option in the GLP Set Menu is turned On, GLP automatically prints data after calibration is completed.
- - - - - SPAN CAL - - - - - -
12/01/99
1:00:00 PM
Bal Id 1234
Cal:
600.0g
Old:
600.0g
Dif:
0.0g
Wt.Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - -
LIN CAL - - - - - -
12/01/99
1:00:00 PM
Bal Id 1234
Cal:
600.0g
Old:
599.9g
Dif:
0.1g
Wt.Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
Span Calibration Printout
When performing Span calibration with all GLP options
turned on, a printout is automatically made after the
calibration is completed.
Linearity Calibration Printout
When performing a Linearity calibration with all GLP
options turned on, a printout is automatically made after
the calibration is completed.