1 configure sensor – VEGA VVO VEGA Visual Operating User Manual
Page 12
12
VEGA Visual Operating
Now click to
Measurement loop, then to
Modify.
Set-up of a process pressure transmitter VEGABAR 20
5 Set-up of a process pressure transmitter VEGABAR 20
5.1 Configure sensor
Connect the sensor by means of
VEGACONNECT 2 to the PC (as de-
scribed under chapter "3 Instrument
connection“). Then start VVO. When the
sensor was connected with already
started VVO, push the function key
F8
on the PC. Then the monitor should
display this picture. To configure the
sensor, first click to
Configuration, then
to
Measuring system.
In this window you can adjust the menu lan-
guage of VEGABAR (only on VEGABAR with
electronics version "HL“). For the meas. unit
you can choose bar, psi or kPa. Click to
Save
after you have modified the adjustments.
By pushing Reset all modifications you have
carried out on VEGABAR (e.g. adjustment)
are reset to factory setting.
With
Quit you return to the window VEGA
Visual Operating.