VEGA Visual Operating

43

Set-up of the processing system VEGALOG

11

Set-up of the processing system VEGALOG

Connect the CPU-card of VEGALOG
and the PC with the RS 232-interface
cable (as described under "3.5 VVO on
the CPU-card“). Then start VVO. When
VVO is already started and you con-
nect VEGALOG, push the function key
F8 on the PC. Then the monitor shows
this picture. First click to

Configuration

and then to

Measuring system to con-

figure VEGALOG.

Your VEGALOG is now displayed ac-
cording to your configuration. By click-
ing to a card (e.g. CPU-card) or
clicking the buttons

<< and >> you

can enquire card-info on all available
VEGALOG-cards. If you have clicked
to the CPU-card (as shown), you can
carry out

Restart (corresponds to

short-term switching off the CPU) or
Reset. With Reset all already created
measurement loops of VEGALOG are
deleted.

When you have enquired EV-card (EV
stands for input - VBUS-sensor), you
can click to the button

Sensor survey.

Then all VBUS-sensors will be listed in
the window "Sensor survey Card-No. X“
which are connected to the EV-card.