Set-up of the processing system VEGALOG

From this window you can also de-
lete sensor data, modify sensor
designations (Sensor-TAG) and
enter serial numbers manually. Click
to the appropriate sensor and then
click to

Configure.

On the AR-card (AR stands for relay
output) and on the AT-card (AT stands
for transistor output) you can carry out
pre-adjustments as on the EV-card.
Click to

Configuration.

In this position a fault signal function
can be coordinated to the switching
outputs (relays or transistors) which
will be later available for creating
measurement loops. The configura-
tion as system failure is a special
feature. Only one switching output
can be defined as system failure
indication. This output does not show
failures of individual measurement
loops but failures of the CPU (e.g. a
total power loss of the CPU).