2 ph – Xylem HandyLab 680 User Manual

Page 106

Advertising
background image

What to do if...

HandyLab 680

106

ba77041e01

03/2014

6.2

pH

Error message,

OFL, UFL

Error message,

Error

Cause

Remedy

IDS pH sensor:

– Measured value outside the

measuring range

– Use suitable IDS pH sensor

– Air bubble in front of the junction

– Remove air bubble

– Air in the junction

– Extract air or moisten junc-

tion

– Gel electrolyte dried out

– Exchange IDS pH sensor

Cause

Remedy

IDS pH sensor:

– The values determined for zero

point and slope of the IDS pH
sensor are outside the allowed
limits.

– Recalibrate

– Junction contaminated

– Clean junction

– IDS pH sensor broken

– Exchange IDS pH sensor

– Buffer solutions:

– The used buffer solutions are not

suitable for the selected buffer
set

– Select a different buffer set

or
– Use different buffer solu-

tions

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change solutions