Xylem HandyLab 680 User Manual

Page 54

Advertising
background image

Operation

HandyLab 680

54

ba77041e01

03/2014

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.

7

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <MENU/ENTER>.
The pH value of the buffer solution is displayed.

8

Set the nominal buffer value for the measured temperature with
<

><>.

9

Accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).

10

If necessary, finish the calibration procedure as a single-point
calibration with <M>.
The calibration record is displayed.

15.03.2014 08:00

VariCal

15.03.2014 08:00

VariCal