HandyLab 680

82

ba77041e01

03/2014

Type

Measuring cell used

Cal

Measuring cells whose cell con-
stant is determined by calibration
in the KCL control standard solu-
tion.
Calibration range:
0.450 to 0.500 cm

-1

The currently valid cell constant
is displayed in the status line.

man

Freely adjustable cell constant in
the range 0.450 ... 0.500 cm

-1

.

Man. cell const.

0.450 ...
0.475 ...
0.500 cm

-1

The cell constant and be dis-
played and set here

Temp. comp. (TC)
/ Method

nLF
Lin
off

Procedure for temperature com-
pensation (see section 4.10.3).

This setting is only available for
the measured parameters,

ϰ and

ρ

.

Temp. comp. (TC)
/
Linear coeff.

0.000 ...
2.000 ...
10.000 %/K

Coefficient of the linear tempera-
ture compensation.

This setting is only available
when the linear temperature
compensation is set.

Temp. comp. (TC)
/ Reference temp.

20 °C
25 °C

Reference temperature

This setting is only available for
the measured parameters,

ϰ and

ρ

.

TDS factor

0.40 ... 1.00

Factor for TDS value

Reset

-

Resets all sensor settings to the
delivery condition (see section
4.14.1)