HandyLab 680

Operation

83

ba77041e01

03/2014

Setting menu of

LF 313T IDS

Menu item

Possible set-
ting

Description

Cell constant

0.090 ... ...
0.100 ...
0.110 cm

-1

The cell constant can be dis-
played and set here

Temp. comp. (TC) /
Method

nLF
Lin
off

Procedure for temperature
compensation (see section
4.10.3).

This setting is only available
for the measured parame-
ters,

ϰ and

ρ

.

Temp. comp. (TC) /
Linear coeff.

0.000 ...
2.000 ...
10.000 %/K

Coefficient of the linear tem-
perature compensation.

This setting is only available
when the linear temperature
compensation is set.

Temp. comp. (TC) /
Reference temp.

20 °C
25 °C

Reference temperature

This setting is only available
for the measured parame-
ters,

ϰ and

ρ

.

TDS factor

0.40 ... 1.00

Factor for TDS value

Reset

-

Resets all sensor settings to
the delivery condition (see
section 4.14.1)