4.10.3 Temperature compensation

The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
Tr20 or Tr25.

You can select one of the following temperature compensation meth-
ods:

Nonlinear temperature compensation (nLF) according to EN 27

888

Linear temperature compensation (Lin) with adjustable coeffi-

cients of 0.000 ... 10.000 %

/

K

 No temperature compensation (off)

Note
The reference temperature and temperature compensation are set in
the menu for the parameter, conductivity (see section 4.10.4).

Application tips

Select the following temperature compensations given in the table
according to the respective test sample:

Test sample

Temperature compensation

Display
indicator

Natural water
(ground water, surface
water, drinking water)

nLF
according to EN 27 888

nLF

Ultrapure water

nLF
according to EN 27 888

nLF

Other aqueous solu-
tions

lin Set linear
temperature coefficient
0.001 ... 10.000 %/K

lin

Salinity (seawater)

Automatic nLF according to
IOT (International Oceanogra-
phic Tables

Sal, nLF