5 calibration interval, Calibration interval – Xylem HandyLab 680 User Manual

Page 48

Advertising
background image

Operation

HandyLab 680

48

ba77041e01

03/2014

Note
For pH IDS sensors, you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 4.7.10).

Preparatory activities

Perform the following preparatory activities when you want to calibrate:

4.7.5

Calibration interval

The calibration evaluation is displayed as a sensor symbol.

After the QSC function has been enabled the sensor symbol is
replaced by the QSC color scale (see section 4.7.10).

After the specified calibration interval has expired, the sensor symbol
or the QSC color scale flashes. It is still possible to measure.

Note
To ensure the high measuring accuracy of the measuring system, cal-
ibrate after the calibration interval has expired.

Setting the calibration

interval

The calibration interval is set to 7 days (d7) in the factory.
You can change the interval (1 ... 999 days):

Error

Error

< -30 or
> 30

... -62 or
... -50

Eliminate the error according to chapter
6 W

HAT

TO

DO

IF

...

Display

Calibration
record

Zero point
[mV]

Slope
[mV/pH]

1

Connect the IDS pH sensor to the meter.
The pH measuring window is displayed.

2

Keep the buffer solutions ready.

1

Open the menu for measurement settings with <MENU/
ENTER>
.

2

In the Calibration / Calibration interval menu, set the calibration
interval with <

><>.

3

Confirm the setting with <MENU/ENTER>.

Advertising