4 ph calibration, Ph calibration – Xylem HandyLab 680 User Manual

Page 43

Advertising
background image

HandyLab 680

Operation

43

ba77041e01

03/2014

4.7.4

pH calibration

Why calibrate?

During the operation of an IDS pH sensor, the zero point (asymmetry)
and slope of the sensor change with time.As a result, an inexact mea-
sured value is displayed. Calibration determines the current values of
the zero point and slope of the pH sensor and stores them in the mea-
suring instrument.
Thus, you should calibrate at regular intervals.

When do you have to

calibrate?

 When the calibration interval has expired

Buffer sets for

calibration

You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into consideration
during the calibration.

Alternative temperature

on
off

Takes the measured temperature value from
an IDS sensor.

This menu item is only available if an IDS
sonsor without temperature sensor and an
IDS sensor with temperature sensor are
connected.

pH resolution

0.001
0.01
0.1

Resolution of the pH display

mV resolution

0.1
1

Resolution of the mV display

Reset

-

Resets all sensor settings to the delivery
condition (see section 4.14.1)

Menu item

Possible setting

Description