Xylem HandyLab 680 User Manual

Page 55

Advertising
background image

HandyLab 680

Operation

55

ba77041e01

03/2014

Continuing with two-

point calibration

11

Thoroughly rinse the IDS sensor with deionized water.

12

Immerse the IDS pH sensor in buffer solution 2.

13

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

><>.

14

Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.

15

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <MENU/ENTER>.
The pH value of the buffer solution is displayed.

16

Set the nominal buffer value for the measured temperature with
<

><>.

17

Accept the calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage dis-
play).

18

If necessary, finish the calibration procedure as a two-point cal-
ibration with <M>.
The calibration record is displayed.

15.03.2014 08:00

VariCal

Advertising