Xylem HandyLab 680 User Manual

Page 74

Advertising
background image

Operation

HandyLab 680

74

ba77041e01

03/2014

Calibration evaluation of

the FDO

®

1100 IDS

Calibration in

water vapor-

saturated air

Proceed as follows to calibrate the D.O. sensor:

Note
The sponge in the check and storage vessel must be moist (not wet).
Leave the sensor in the check and storage vessel long enough so it can
adapt to the ambient temperature.

Display

Calibration record

Relative slope

+++

S = 0.94 ... 1.06

++

S = 0.92 ... 0.94 or
S = 1.06 ... 1.08

+

S = 0.90 ... 0.92 or
S = 1.08 ... 1.10

Error

Error

S < 0.90 or
S > 1.10

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

1

Connect the D.O. sensor to the meter.

2

Place the

FDO

®

1100 IDS D.O. sensor into the check and stor-

age vessel.

3

Start the calibration with <CAL>.
The last calibration data (relative slope) is displayed.