HandyLab 680

Operation

67

ba77041e01

03/2014

Note
If the HOLD function is active, you can, e.g. start a manual measure-
ment with stability control.

Stability control

(AutoRead )

The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values. The display of the measured
parameter flashes until a stable measured value is available.

You can start a measurement with Stability control manually at any
time, irrespective of the setting for automatic Stability control (see sec-
tion 4.4.3
) in the System menu.

Note
You can prematurely terminate the Stability control function manually
with <MENU/ENTER> at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are not output to
the interface.

2

Release the frozen measured value again with <AR>.
The HOLD function is switched off.
The [HOLD] status display disappears.

1

Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.

2

Using <MENU/ENTER>, activate the Stability control function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked
by AR.

3

Using <MENU/ENTER>, start a further measurement with Sta-
bility control
.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.