2 firmware-update for ids sensors, Firmware-update for ids sensors – Xylem HandyLab 680 User Manual

Page 116

Advertising
background image

Appendix: Firmware update

HandyLab 680

116

ba77041e01

03/2014

has taken over the new software version (see page 103).

9.2

Firmware-Update for IDS Sensors

General information

Available firmware updates for IDS sensors are provided on the Inter-
net. With the "Firmware Update" program and a PC you can update the
firmware of an IDS sensor to the newest version.

For updating, connect the IDS sensor to the HandyLab 680, and the
HandyLab 680 to a PC.

For the update via the USB-B interface, the following is required:

 a free USB interface (virtual COM port) on the PC

 the driver for the USB interface (on the enclosed CD-ROM)

 the USB cable (included in the scope of delivery of the

HandyLab 680).

Program installation

Program start

Firmware update

1

Install the downloaded firmware update on a PC.

An update folder is created in the Windows start menu.
If an update folder already exists for the sensor (or sensor
type), the new data are displayed there.

2

In the windows start menu, open the update folder and start the
firmware update program for the IDS sensor

3

Connect the IDS sensor to the meter.
Only the sensor connection in the lower section of the socket
field (channel 1) is suitable for firmware update.

4

Using the USB interface cable, connect the HandyLab 680 to a
USB interface (virtual COM port) of the PC.

5

Switch on the HandyLab 680.

6

In the firmware update program, start the update process with
OK.

7

Follow the instructions of the firmware update program.
During the programming process, a corresponding message
and a progress bar (in %) are displayed.
The programming process takes up to 5 minutes. A terminatory
message is displayed after a successful programming process.
The firmware update is completed.

8

Disconnect the HandyLab 680 from the PC.
Meter and sensor are ready for operation again.