Operation

HandyLab 680

46

ba77041e01

03/2014

*

Brand names or trade names are trademarks of their respective
owners protected by law.

Note
The buffers are selected in the menu, pH / <MENU/ENTER> / Calibra-
tion
/ Buffer (see section 4.7.3).

Calibration points

Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:

Note
You can display the slope in the units, mV/pH or % (see section 4.7.3).

Stability control

The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be termi-
nated at any time (accepting the current value).

Calibration record

The new calibration values are displayed when a calibration procedure

23

Chemsolute *

2.000
4.000
7.000
10.000

20 °C

24

USABlueBook

*

4.000
7.000
10.000

25 °C

25

YSI *

4.000
7.000
10.000

25 °C

No.

Buffer set*

pH values

at

Determined
values

Displayed calibration data

1-point

Asy

 Zero point = Asy
 Slope = Nernst slope

(-59.2 mV/pH at 25 °C)

2-point

Asy
Slp.

 Zero point = Asy
 Slope = Slp.

3-point to
5-point

Asy
Slp.

 Zero point = Asy
 Slope = Slp.

The calibration line is calculated
by linear regression.