Xylem HandyLab 680 User Manual

Page 47

Advertising
background image

HandyLab 680

Operation

47

ba77041e01

03/2014

is completed.

Display calibration data

and output to interface

You can have the data of the last calibration displayed (see section
4.7.8
). Subsequently, you can transmit the displayed calibration data to
the interface, e. g. to a PC, with the <PRT> key.

Note
The calibration record is automatically transmitted to the interface after
calibrating.

Sample record

Calibration evaluation

After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.

HandyLab 680
Ser. no. 09250023

CALIBRATION

pH

Calibration date

15.03.2014 16:13:33

BlueLine 24 pH IDS
Ser. no. B092500013

TEC
Buffer 1

4.01

Buffer 2

7.00

Buffer 3

10.01

Voltage

1

184.0 mV 24.0 °C

Voltage

2

3.0 mV

24.0 °C

Voltage 3

-177.0 mV 24.0 °C

Slope

-60.2 mV/pH

Asymmetry

4.0 mV

Sensor

+++

etc...

Display

Calibration
record

Zero point
[mV]

Slope
[mV/pH]

+++

-15 ... +15

-60.5 ... -58

++

-20 ... +20

-58 ... -57

+

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-

-30 ... +30

-62 ... -61
or
-56 ... -50

Clean the IDS sensor according to the
sensor operating manual