5 d.o. calibration, D.o. calibration – Xylem HandyLab 680 User Manual

Page 73

Advertising
background image

HandyLab 680

Operation

73

ba77041e01

03/2014

4.9.5

D.O. calibration

Why calibrate?

D.O. sensors age. This changes the slope of the D.O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
the instrument.

Note
The

FDO

®

1100 IDS D.O. sensor ages so little it does not have to be

regularly calibrated.
To detect changes of the sensor as early as possible, the FDO

®

Check

procedure can be useful (see section 4.9.4).

When to calibrate?

 If your evaluation of the FDO

®

Check procedure suggests a calibra-

tion

 When the calibration interval has expired

 If your accuracy requirements are especially high

 Routinely within the scope of company quality assurance measures.

Calibration procedures

Calibration in water vapor-saturated air.
To calibrate the

FDO

®

1100 IDS sensor, use the check and storage

vessel.

Stability control

(AutoRead)

In calibration, the Stability control function (AutoRead) is automatically
activated.

Display calibration data

and output to interface

You can have the data of the last calibration displayed (see section
4.10.6). Sub
sequently, you can transmit the displayed calibration data
to the interface, e. g. to a PC, with the <PRT> key.

Note
The calibration record is automatically transmitted to the interface after
calibrating.

Sample record

Calibration evaluation

After the calibration, the meter automatically evaluates the current sta-
tus of the calibration. The evaluation appears on the display and in the
calibration record.

HandyLab 680
Ser. no. 10139695

CALIBRATION Ox
Calibration date

15.03.2014 16:13:33

FDO 1100 IDS
Ser. no. 10146858

OX 930

10158765

Relative slope

0.98

Sensor

+++

Advertising