HandyLab 680

68

ba77041e01

03/2014

Criteria for a stable

measured value

The Stability control function checks whether the measured values are
stable within the monitored time interval.

The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.

4.9.3

Settings for D.O. sensors
(menu or measurement and calibration settings)

Overview

The following settings are possible for D.O. sensors:

 Calibration record (display)

 Calibration data storage (display)

 Calibration interval

 FDO Check

 Check interval

 Salinity correction

 Salinity (salinity equivalent)

Settings

The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the <MENU/ENTER> key. After
completing the settings, switch to the measured value display with
<M>.

Measured
parameter

Time interval

Stability in the time interval

D.O. concen-
tration

20 seconds

Δ

: Better than 0.03 mg/l

D.O. saturation

20 seconds

Δ

: Better than 0.4 %

D.O. partial
pressure

20 seconds

Δ

: Better than 0.8 mbar

Temperature

15 seconds

Δ

: Better than 0.5 °C

Menu item

Possible
setting

Description

Calibration /Cali-
bration record

-

Displays the calibration record of the
last calibration