Setting the measuring mode

WQL-Cond

124

ba75822e01

03/2011

5

Setting the measuring mode

You can make the following settings in the Device / Set menu (default setting
in bold):

TDS factor

The factor to calculate the total dissolved solids is set to 1.00 in the factory.
You can adjust this factor to meet your requirements in the range of 0.40 to
1.00. The factor is set in the menu for the parameter, TDS.

Temperature

compensation

The calculation of the temperature compensation is based on the preset
reference temperature, 20 °C or 25 °C. The reference temperature is shown
on the display.

-Temperature

coefficient

The temperature coefficient for the linear temperature compensation is set to
2.000 %/K in the factory. You can adjust this factor to meet your requirements
in the range 0.000 ... 3.000 %/K. The factor is set in the menu for the setting
of the temperature compensation.

Note
More information on conductivity measurements is given in the WTW
conductivity primer.

Measuring mode

z

Conductivity
[

μS/cm] / [mS/cm]

z

Salinity Sal [ ]

z

TDS
[mg/l] / [g/l] with adjustable TDS factor
in the range 0.40 ... 1.00

z

Resistivity
[

Ω·cm] / [kΩ·cm] / [MΩ·cm]

Temperature compensation

z

Nonlinear temperature
compensation
(nLF) according to EN
27 888

z

Linear temperature compensation (lin)
with adjustable coefficients in the
range 0.000 ... 2.000 ... 3.000 %

/

K

z

No temperature compensation (off)

Reference temperature

z

25 °C

z

20°C