Xylem WQL-Cond CONDUCTIVITY DATALOGGERS User Manual

Page 46

Advertising
background image

Logging

WQL-Cond

142

ba75822e01

03/2011

Temperature

compensation

The calculation of the temperature compensation is based on the preset
reference temperature, 20 °C or 25 °C. The reference temperature is shown
on the display.

Temperature

coefficient

The temperature coefficient for the linear temperature compensation is set to
2.000 %/K in the factory. You can adjust this factor to meet your requirements
in the range 0.000 ... 3.000 %/K. The factor is set in the menu for the setting
of the temperature compensation.

TDS factor

The factor to calculate the total dissolved solids is set to 1.00 in the factory.
You can adjust this factor to meet your requirements in the range of 0.40 to
1.00. The factor is set in the menu for the parameter, TDS.

Time

Automatic time
adjustment

z

Yes (date/time for the logger is
automatically synchronized with that
of the PC)

z

No
(Enter the date/time for the logger)

Parameter

Mode

z

Conductivity
[

μS/cm] / [mS/cm]

z

Salinity Sal [ ]

z

TDS
[mg/l] / [g/l] with adjustable TDS
factor in the range 0.40 ... 1.00

z

Resistivity
[

Ω·cm] / [kΩ·cm] / [MΩ·cm]

Temperature
compensation

z

Nonlinear temperature
compensation
(nLF) according to
EN 27 888

z

Linear temperature compensation
(lin) with adjustable coefficients in
the range 0.000 ... 2.000 ... 3.000 %/
K

z

No temperature compensation (off)

Reference
temperature

z

25 °C

z

20°C

Tab

Setting

Setting options
(default setting in bold)