13.2 Communication of the logger and PC program

Import of

measured values

is aborted

Data logging of the

logger aborted

13.3 Communication between PC program and SQL server

The program

cannot access the

server database

Cause

Remedy

– Communication problem

– Repeat the import process

– Use a shorter USB cable

– USB-Hub or USB connection at

the PC faulty or defective

– Use other USB connection or PC

Cause

Remedy

– Power supply interrupted

– Connect the logger to the PC

program and set up logging job.

– Make sure that the logger is

supplied with power (battery)

Cause

Remedy

– The SQL server is already

installed on the PC with a
different password

– In the Settings / Database menu,

click the User-defined option and
enter the password that was used
for the initial installation of the SQL
server.

– Set Windows authentication.

Note
The connection can be checked with the Test button. A message appears
informing you whether the connection was established.