Xylem WQL-Cond CONDUCTIVITY DATALOGGERS User Manual

Page 29

Advertising
background image

WQL-Cond

Setting the measuring mode

125

ba75822e01

03/2011

Application tips

Select the temperature compensations given in the table according to the
respective test sample:

Test sample

Temperature compensation

Display

Natural water
(ground water, surface
water, drinking water,
waste water)

nLF
according to EN 27 888

nLF

Other aqueous solutions

Set linear
temperature coefficient
0.000 ... 3.000 %/K

Linear

Salinity (seawater)

Automatic according to IOT
(International Oceanographic
Tables)

1

In the Device menu, select the menu item, Set.

2

The window for the selection of the measuring mode, temperature
compensation and reference temperature appears.

3

Select the required settings with a mouse click.
Clic Send settings.
A confirmation message appears: