WQL-Cond

Calibration in control standard (determining the cell constant)

129

ba75822e01

03/2011

6.2

Calibrating (automatic)

Preparatory

activities

Calibration

Note
We recommend to always calibrate with the protective hood screwed on so
the measuring cell is protected. Use a stand as necessary.

1

Connect the WQL-Cond logger to a USB port of your PC with the
USB cable.

2

Have the 0.01 mol/l KCl control standard ready.

3

In the Device menu, click the menu item, Calibrate.
The calibration window appears.

4

Immerse the logger in the 0.01 mol/l KCl control standard.

5

Use the Next button to start the measurement.