Measuring directly

The WQL-Log PC program in conjunction with the WQL-Cond logger can
measure and display the value of a solution directly according to the set
parameter. The directly measured values are not stored in the logger.

Proceed as follows when you want to measure directly:

1

Connect the WQL-Cond logger to the PC with the USB cable.

2

In the Device/Login menu, connect the logger to the WQL-Log PC
program.

3

If a logging job is running, terminate or interrupt it (measuring directly
is not possible while a logging job is running).

4

In the Device menu, click the Set menu item and select the settings
for the measuring mode and temperature compensation (see chapter
5 S

ETTING

THE

MEASURING

MODE

).

5

In the Device menu, click the Measure directly menu item.

6

Immerse the WQL-Cond logger in the measuring solution.

7

The measured value window pops up with the measured value
(depending on the setting) and the temperature value of the test
sample.
Next to the measured value, the set temperature compensation and
if necessary, the temperature coefficient or TDS factor and reference
temperature are displayed.

8

Terminate the direct measurement with Quit.