Calibration in control standard (determining the cell constant)

WQL-Cond

134

ba75822e01

03/2011

Calibration record

7

Enter the conductivity value for the displayed temperature.

8

Click Continue. As soon as a stable value is recognized, the
calibration record appears.

9

Accept the calibration with Accept. The message Calibration
successful
appears and the new calibration will now be used for
measurement.
If you press Cancel, the logger discards the new calibration and will
continue to use the previous calibration values.