WQL-Cond

Resetting the logger

177

ba75822e01

03/2011

11

Resetting the logger

You can reset to the default condition the logger and the PC program with its
settings.

In the default condition, the settings are as follows:

Resetting to

default condition

To reset the logger and PC program to the default condition, proceed as
follows:

Configuration parameters

Setting

Logging interval

1 second

Start of the logging

Logger key button

End of the logging

Time period
1 hour

Measured value ID

default

Measuring mode

Conductivity

Temperature compensation

nLF

Temperature coefficient

2.000

Reference temperature

25 °C

Cell constant

0.475 cm

-1

TDS factor

1.00

Date, time

01.01.2003, 00:00

Calibration records

None

Notes
z

The Device / Reset function deletes the calibration records as well as all
measurement data from the logger! Before deleting the logger memory,
make sure that you have imported to the database all data you want to
keep.

z

The battery counter is not reset with the Device / Reset function. It has to
be reset with the reset function of the battery manager.

1

In the Device menu, click the menu item, Reset. A security prompt
appears.

2

Confirm with OK. The settings are reset to the default condition.